Z-0477-2021: Verigene Processor SP, Model Numbers 10-0000-07 and 10-0000-07R. Used with assay Luminex, VERIGENE Gram-

Recall record Z-0477-2021 for Luminex Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0477-2021
Company in FDA record
Luminex Corporation
FDA record creation date
Nov 20, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Luminex has determined that three (3) separate test fixtures used by field service engineers to service the VERIGENE SP Processors were procured from unapproved sources and, in some cases, were manufactured using unvalidated processes. These text fixtures were used to calibrate and verify VERIGENE SP devices.

Source facts

Action
On 10/13/2020, the firm sent the following notice: "As part of our corrective action plan, Luminex will be updating software on all VERIGENE SP instruments in the coming weeks, and systematically visiting every VERIGENE customer site to perform a supplemental preventative maintenance on your instruments. We will also likely be sending out a series of Customer Advisory Notifications with additional information on items related to our remediation efforts surrounding the Warning Letter, as well as updates related to other items identified for improved product performance. During this time, you may continue to use your VERIGENE products as directed in the package inserts for each product, or as otherwise directed through further Luminex communications. If you have any question, please do not hesitate to contact Luminex Global Support Services using the contact details below: Luminex Global Support Services 1-877-785-2323 (U.S. and Canada) +1-512-381-4397 (Outside U.S. and Canada) support@luminexcorp.com"
Initiated
2020-10-13
Root cause
Nonconforming Material/Component
Terminated
2024-01-22
Distribution
Worldwide distribution. US nationwide and Kuwait, Austria, United Kingdom, Spain, France, Chile, Japan, Netherlands, Greece, Italy, Thailand, Germany, Turkey
Recall number
Z-0477-2021
Recall status
Terminated
CDRH record ID
184502
FDA record created
2020-11-20
Recall event number
86124

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗