Skip to content

Public FDA record

Z-0477-2022: OCT-Camera 211 01 A1, Running software iOCT-Control 3 v1.4

Recall record Z-0477-2022 for Haag-Streit USA Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0477-2022
Company in FDA record
Haag-Streit USA Inc
FDA record creation date
Jan 10, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Malfunction of the automatic laser beam shut-off, the OCT unit might not recognize whether the laser beam is safely switched off

Source facts

Action
Haag-Streit USA send out field safety notice 2021-10/01 on November 03, 2021. Actions to be taken by the customer/user: Should one or more continuous straight lines be displayed on the screen of the OCT camera during an operation on the posterior segment of the patient's eye (see picture in the notification), the described risk possibly exists. 1. Switch off the OCT camera immediately (menu OCT -> Switch off). 2. The OCT camera is not permitted to be used any more. Please contact your responsible country representative. 3. Perform the software update of the OCT camera. You will receive the new software version via your local representative, who will correct the error. 4. Please fill out and return the enclosed response form as soon as the software update has been performed 5. Please ensure that all users of the products mentioned are aware of and receive this urgent safety information. If you have passed the products to third parties, please forward a copy of this information and inform the contact person listed in the notification. 6. Software update organization will be performed by Haag-Streit USA, Inc. Please contact Mike Dietz at (513) 486-1513 or via email: fieldsafetynotice@haag-streit-usa.com for questions concerning this procedure.
Initiated
2021-11-03
Root cause
Software design (manufacturing process)
Distribution
US Nationwide distribution in the states of OH, IN.
Recall number
Z-0477-2022
Recall status
Open, Classified
CDRH record ID
190715
FDA record created
2022-01-10
Recall event number
89176

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.