Skip to content

Public FDA record

Z-0478-06: CryoValve, Aortic Valve &Conduit

Recall record Z-0478-06 for Cryolife Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0478-06
Company in FDA record
Cryolife Inc
FDA classification date
Feb 4, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

During a retrospective review, CryoLife identified positive microbiological culture results for donor tissues associated with various allografts.

Source facts

Action
Consignees were notified by letter dated May 6, May 11, or May 12, 2004.
Initiated
2004-05-06
Root cause
Other
Terminated
2010-06-28
Distribution
FL, NC , NE, NV, NY, TN, TX, UT, WA, and Canada
Recall number
Z-0478-06
Recall status
Terminated
CDRH record ID
35517
FDA classified
2006-02-04
Recall event number
30233

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗