The hanger bar assembly may detach from the lift during use due to fatigue failure of the bolt.
Source facts
Action
Arjo sent an Urgent Device Field Correction Recall letter dated 8/31/06 to all affected end users to the attention of the Administrator/Risk Manager via FedEx on the same date. The letters informed the accounts of the potential for the hanger bar to detach from the scale due to fatigue failure of the pivot bolt. The accounts were advised that an Arjo Field Service Technician would contact them within five business days to schedule a visit to the account to perform an upgrade of their Minstrel lifts. Any questions were directed to Arjo Quality Department at 800-323-1245, ext. 6140.
Initiated
2006-08-31
Root cause
Other
Terminated
2008-10-16
Distribution
Nationwide
Recall number
Z-0478-2007
Recall status
Terminated
CDRH record ID
48151
FDA classified
2007-01-11
Recall event number
36187