Z-0478-2009: Micro Plate EIA Oral Fluid Negative Control which is a component of the Intercept Micro Plate EIA kit.

Recall record Z-0478-2009 for OraSure Technologies, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0478-2009
Company in FDA record
OraSure Technologies, Inc.
FDA classification date
Dec 23, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Results, false-positive test: low % displacement result of 36.0% (spec 39-60%). For further information, please contact the firm at 610-882-1820.

Source facts

Action
The recalling firm issued a letter entitled "URGENT PRODUCT RECALL NOTICE" dated 9/8/08 to their customers to inform them of the problem and the need to return the product.
Initiated
2008-09-08
Root cause
Nonconforming Material/Component
Terminated
2009-05-05
Distribution
Worldwide Distribution --- USA including states of AZ, CA, CT, DE, FL, GA, ID, KS, KY, MA, MD, MI, NC, NH, NV, NY, PA, VA, and WY, and countries of United Kingdom, Italy, Germany, and Canada.
Recall number
Z-0478-2009
Recall status
Terminated
CDRH record ID
73807
FDA classified
2008-12-23
Recall event number
49557

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗