Z-0478-2021: RayStation standalone software treatment planning system, Model No. 4.0, 4.5, 4.7, 4.9, 5.0, 6.0, 6.1, 6.

Recall record Z-0478-2021 for RAYSEARCH LABORATORIES AB. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0478-2021
Company in FDA record
RAYSEARCH LABORATORIES AB
FDA record creation date
Nov 20, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

If a region of interest (ROI) or point of interest (POI) that is referenced from an imported plan is missing in the imported RT Structure Set, the reference may become linked to the wrong ROI or POI.

Source facts

Action
On October 8, 2020, the firm emailed "Field Safety Notice, Medical Device Correction #66704" to affected customers. Customers were made aware of the product issue and asked to take the following actions: " Do not exclude the prescription ROI or POI when exporting from RayStation with the Exclude from export option. " If an RT Structure Set where it is not certain that all referenced bolus and prescription ROIs or POIs are present is imported to RayStation, make sure to check that any intended references to bolus and prescription are correct after import. " Please educate planning staff and all users about this workaround. " Inspect your product and identify all installed units with the above software version number(s). " Confirm you have read and understood this notice by replying to the notification email. This issue will be resolved in the next version of RayStation, scheduled for market release in November 2020 (subject to market clearance in some markets). If customers wish to continue using versions of RayStation affected by this notice, all users must maintain awareness of this notice. Alternatively, customers can choose to upgrade to the new version once it becomes available for clinical use. For regulatory information, please contact quality@raysearchlabs.com
Initiated
2020-10-08
Root cause
Software design
Terminated
2023-03-30
Distribution
US Nationwide distribution.
Recall number
Z-0478-2021
Recall status
Terminated
CDRH record ID
184282
FDA record created
2020-11-20
Recall event number
86668

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗