Z-0479-2022: Software BeamAdjust version 2.2, Catalog No: S070009 Software VeriSoft version 8.0, Catalog No: S080032

Recall record Z-0479-2022 for PTW NORTH AMERICA CORPORATION. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0479-2022
Company in FDA record
PTW NORTH AMERICA CORPORATION
FDA record creation date
Jan 11, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

When a measurement with a PTW detector array is performed with the software BeamAdjust 2.2 or VeriSoft 8.0, a measurement error can occur under specific conditions. This error can lead to a too-low or too-high measured absolute dose which affects patient plan verification results

Source facts

Action
PTW US notified customer Urgent Safety Notice via email and telephone on 8/27/20. Letter states reason for recall, health risk and action to take: PTW will provide free-of-charge software updates. Users who use BeamAdjust 2.2/ VeriSoft 8.0 according to the above described application must update the software to VeriSoft 8.0.1/BeamAdjust 2.2.1 prior to further use of the software. For all other users PTW recommends using BeamAdjust 2.2.1/VeriSoft 8.0.1 too. Until the software updates are available PTW recommends avoiding the air density correction method. As a workaround, if air density correction should still be applied a manually calculated air density correction value can be entered as a kUser keeping kTP untouched (kTP =1.000 !). Please also note, machine QA measurements performed with PTW detector arrays within the PTW software MultiCheck are not affected. TW has fixed the bug in the software and provides all affected users with software updates.
Initiated
2021-08-27
Root cause
Software design
Terminated
2024-02-09
Distribution
US Nationwide distribution in the state of CO.
Recall number
Z-0479-2022
Recall status
Terminated
CDRH record ID
191251
FDA record created
2022-01-11
Recall event number
89314

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗