Skip to content

Public FDA record

Z-0479-2026: BEAR Implant (Bridge-Enhanced ACL Restoration); Model/Catalog: 1000;

Recall record Z-0479-2026 for Miach Orthopaedics. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0479-2026
Company in FDA record
Miach Orthopaedics
FDA classification date
Nov 7, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The field action was initiated in February of 2023 and was limited to the affected lot from sales force trunk stock only.

Source facts

Action
On January 30, 2023 RE: BEAR Implant Hydration memo emails were sent to sales staff. If you observe an implant that hydrates too quickly (e.g., less than 30 seconds) and appears that it may not remain intact upon insertion into the knee, do not attempt to implant the device. We would recommend you always have a backup device available to address such circumstances. We plan to revise the Tips and Pearls document, ML-1027, to provide further instructions/details on how to properly hydrate the BEAR device. In the meantime, one way to avoid shortened hydration time is to consider hydrating the dense side of the implant first and continue toward the porous side . To date, all reports of this observation have involved Lot # 7008409. As a precaution and to ensure customer satisfaction, we are asking that you return all devices from this lot that are currently in your trunk stock (in your possession). Chris Brown will work with you to coordinate returns as well as replacing these trunk stock devices. Please contact me (mohara@miachortho.com) or Arjun (aishwar@miachortho.com) if you have questions.
Initiated
2023-02-02
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution.
Recall number
Z-0479-2026
Recall status
Open, Classified
CDRH record ID
216478
FDA classified
2025-11-07
Recall event number
97742

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗