Reason stated in the source
Product monitoring has determined that the connection bracket shows potential mechanical play at the OR table interface for some procedures. This is due to the hole in the lock washer, which is responsible for locking the accessory to the OR table, being too large.
Source facts
- Action
- Urgent Device Correction Letters were sent by Federal Express to affected US hospitals on October 22, 2009. The letter stated the technical description, potential hazard, and the recommended precaution. The letter asked customers to immediately verify whether their Connection Bracket is affected thru the FaxBack provided. Questions or need for additional information can be directed to customers' local MAQUET representative.
- Initiated
- 2009-10-22
- Root cause
- Device Design
- Terminated
- 2010-05-18
- Distribution
- Nationwide Distribution -- USA, including states of SC and PA.
- Recall number
- Z-0480-2010
- Recall status
- Terminated
- CDRH record ID
- 86036
- FDA classified
- 2009-12-02
- Recall event number
- 53612