Out of tolerance for radio frequency emissions. At the 150-1000MHz frequency, the testing indicated the BrightMatter Navigation system was up to 20dB uV/meter higher than the applicable IEC 60601-1-2:2007 (Ed3.0) standard specification.
Source facts
Action
Consignees were sent via courier on 12/4/2015 a Synaptive "Urgent Medical Device Recall" letter dated 30 November 2015. The letter described the problem and the device involved in the recall. Informed consignees that a technical support team member from Synaptive Medical will conduct the field correction. They also requested consignees to complete and return the Recall Response Form. For questions contact Thas Yuwaraj, PhD, .PEng Monday through Friday, 9 AM to 5 PM, Eastern Time at 1-844-462-7246.
Initiated
2015-12-04
Root cause
Component design/selection
Terminated
2016-03-04
Distribution
US: FL, WI, IL, LA, WA.
Recall number
Z-0480-2016
Recall status
Terminated
CDRH record ID
142094
FDA classified
2015-12-23
Recall event number
72747