Skip to content

Public FDA record

Z-0480-2021: EMPOWR Partial Knee Peg Drill; Model: 800-06-008.

Recall record Z-0480-2021 for Encore Medical, LP. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0480-2021
Company in FDA record
Encore Medical, LP
FDA record creation date
Nov 20, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There have been issues experienced with the peg preparation steps, specifically drilling of the Femoral posterior peg. Upon drilling, the shaft portion of the drill may bind within the Posterior Peg drill guide. The extended length of the drill guide and tight tolerances to ensure accurate preparation seem to be contributing to this situation. There is a potential of a delay in surgery if this issue were to recur.

Source facts

Action
The method of notification was via e-mail to surgical sales agents, and the notices were issued on 06/18/2020. Additionally, a notification to return the affected drills for replacement were sent via e-mail to affected surgical agents on 07/20/20. The following recommendations should be communicated to the implanting surgeons: Sales Agents Action: To avoid the potential binding, the following is recommended: " Use a plunge mill approach, i.e. plunge the drill in and out of the drill guide bushing a few times slowly until the hole is fully prepared. " Use of saline or fat to keep the drill lubricated during use. " Inspect all instrumentation prior to use to identify any wear, breakage or damage. Along with the notification, the communication includes an acknowledgment form that should be completed and returned to confirm that the instructions provided have been read and understood. THE ACKNOWLEDGEMENT SHOULD BE REVIEWED AND COMPLETED BY JULY 6, 2020.For questions, contact Teffany Hutto by phone at (512) 834-6255 or by email at teffany.hutto@djoglobal.com.
Initiated
2020-06-19
Root cause
Other
Distribution
US: AZ, CA, FL, GA, IL, IN, MI, MN, NY, TX, VA, and WA.. No OUS Consignees.
Recall number
Z-0480-2021
Recall status
Open, Classified
CDRH record ID
182220
FDA record created
2020-11-20
Recall event number
85943

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗