Skip to content

Public FDA record

Z-0480-2023: PDS II (polydioxanone) Suture- VIOLET MONOFILAMENT 27IN 0 S/A CT-1 Product Code: Z340H For use in soft t

Recall record Z-0480-2023 for Ethicon, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0480-2023
Company in FDA record
Ethicon, Inc.
FDA classification date
Dec 14, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

knot tensile strengths were observed to be significantly below the minimum knot tensile, will result in breakage prior to use or intra-operatively rather than premature post-operative breakage

Source facts

Action
Ethicon notified the four impacted countries via "URGENT: MEDICAL DEVICE RECALL (REMOVAL)" beginning Nov 16, 2022. Letter states reason for recall, health risk and actions to be taken. The customers were instructed to: Examine your inventory immediately to determine if you have product subject to this recall on hand and quarantine such product(s). 2. Remove the product subject to this recall and communicate the issue to relevant operating room or materials management personnel, or anyone else in your facility who needs to be informed. 3. If any product subject to this recall has been forwarded to another facility, contact that facility to arrange return. Please consider including a copy of this recall letter when communicating. 4. Complete the Business Reply Form (BRF) (Attachment 2) confirming receipt of this notice and return to Ethicon, Inc. within three (3) business days. Please return the BRF even if you do not have product subject to this recall. 5. Keep this notice visibly posted for awareness until all product subject to this recall has been returned. While processing your returns, please maintain a copy of this notice with the product subject to this recall and keep a copy for your records. 6. Customers are required to return unused PDS" II products subject to this recall that are in inventory immediately. To receive credit reimbursement, customers must return product subject to this recall to Ethicon, Inc. Any non-affected product and any product returned after the date specified will not receive credit reimbursement. If you have any questions, call 908-218-0707.
Initiated
2022-11-16
Root cause
Under Investigation by firm
Distribution
International (OUS) only: Canada, Japan, Australia and New Zealand
Recall number
Z-0480-2023
Recall status
Open, Classified
CDRH record ID
197225
FDA classified
2022-12-14
Recall event number
91192

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗