Z-0480-2026: GM85 Digital Mobile X-ray imaging System; Model Number: GM85;

Recall record Z-0480-2026 for NeuroLogica Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0480-2026
Company in FDA record
NeuroLogica Corporation
FDA classification date
Nov 7, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The four (4) M5 bolts that secure the Tube Head Unit to the end of the arm may loosen or fracture due to metal fatigue.

Source facts

Action
On October 7, 2025, "URGENT MEDICAL DEVICE CORRECTION" letters were sent to customers. The letters described the problem, product and actions to be taken. The customers were instructed to take the following Actions: Users may continue to use the GM85 system. Customers can inspect the unit for missing or loose bolts and if any issues with the bolts are detected, then please contact the service engineer immediately for inspection. This notice needs to be passed on all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Please transfer this notice to other organizations on which this action has an impact. Please maintain awareness on this notice and resulting action for an appropriate period to ensure effectiveness of the corrective action. Complete and return the Customer Reply Form (CRF) Samsung Electronics will undertake the voluntary corrective action to reinforce the arm with a high tensile bolt to eliminate the risk of the THU falling and preventing any potential risk of injury. Service engineers will correct all affected devices free of charge and contact you to arrange for the correction. All corrections in the field started in September 2025 and will be completed by June 2026. If you have any questions, contact VP via email: ngujar@neurologica.com or call 978-564-8632.
Initiated
2025-10-07
Root cause
Device Design
Distribution
Domestic: US Nationwide Distribution. International: Australia, Belgium, Canada, Chile, China Mainland, Czech Republic, Denmark, Ecuador, Finland, France, Germany, GHANA, Hungary, India, Italy, Korea, Malaysia, Mexico, Netherlands, New Zealand, Norway, Peru, Philippines, Poland, Portugal, Saudi Arabia, Serbia, Singapore, Spain, Sweden, Switzerland, Thailand, T¿rkiye, United Kingdom, Utd.Arab Emir., Vietnam,
Recall number
Z-0480-2026
Recall status
Open, Classified
CDRH record ID
216399
FDA classified
2025-11-07
Recall event number
97721

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗