Open original FDA record ↗Recall record
- FDA identifier
- Z-0482-2007
- Company in FDA record
- Arjo, Inc.
- FDA classification date
- Jan 11, 2007
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The carabineer may not be correctly attached to the reacher and the portable patient lift may drop.
Source facts
- Action
- BHM sent recall letters to Arjo''s direct accounts on 11/13/06 via first class mail, informing them that fastening the carabineer to the exterior of the closed ring portion of the reacher is contrary to the manufacturer''s instructions and is an unsafe practice. The letters informed the users that Health Canada had received notification of an incident involving a reacher marketed by a competitor, and had conducted an investigation into all similar reachers available on the Canadian market. Enclosed with the letter was a copy of the BHM Reacher User''s Guide as well as a warning sticker that visually describes the safe method for connecting the carabineer to the reacher. The accounts were requested to place the warning stickers on their reachers, review the enclosed User''s Guide in detail, and ensure that all personnel using the reacher receive refresher training on its proper use and have access to the supplied User''s Guide. Once these measures have been completed, the account was requested to complete the enclosed response form and return it to BHM by fax, e-mail or mail.
- Initiated
- 2006-11-13
- Root cause
- Other
- Terminated
- 2010-01-19
- Distribution
- Nationwide
- Recall number
- Z-0482-2007
- Recall status
- Terminated
- CDRH record ID
- 49378
- FDA classified
- 2007-01-11
- Recall event number
- 36802