Z-0482-2010: TempSpan Dual-Cure Temporary Crown and Bridge Material, PART NUMBER N69AD, The intended use of this dev

Recall record Z-0482-2010 for Jeneric/Pentron. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0482-2010
Company in FDA record
Jeneric/Pentron
FDA classification date
Dec 3, 2009
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Pentron Clinical Technologies LLC received two (2) complaints alleging that the TempSpan Dual-Cure Temporary Crown and Bridge Material was not setting. Investigation of product in inventory was conducted and it was confirmed that some of the bottles had been mispackaged with an incorrect catalyst.

Source facts

Action
The recall customer notification was initiated on October 22, 2009 with the firm forwarding an urgent medical device letter with recall return form via USPS first class mail to the affected consignees. The letter informed the consignees of the reason for recall and instructed them to return any used or unused product of the affected lot. The firm provided an 800 number for the consignees to call to arrange for product return and replacement.
Initiated
2009-10-22
Root cause
Employee error
Terminated
2012-07-19
Distribution
Nationwide
Recall number
Z-0482-2010
Recall status
Terminated
CDRH record ID
86306
FDA classified
2009-12-03
Recall event number
53673

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗