Open original FDA record ↗Recall record
- FDA identifier
- Z-0482-2011
- Company in FDA record
- Boston Scientific Corporation
- FDA classification date
- Nov 26, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The product may contain manufacturing aids (mandrels) which should not be present in the final product. If the mandrel remains in place, the catheter cannot be loaded over the required guidewire, so cannot be used.
Source facts
- Action
- Boston Scientific issued an Urgent Customer Notice dated March 1, 2010 to all US customers, with personal delivery by sales representative where applicable outside the US, identifying the affected device. Customers were instructed to return the device for replacement, and to return an acknowledgement form. Customers can contact Deborah Springer at Boston Scientific at 800-845-4598 about this recall.
- Initiated
- 2010-03-01
- Root cause
- Process control
- Terminated
- 2010-11-26
- Distribution
- Worldwide Distribution: Throughout the US and to the countries of: Austria, Australia, Belgium, Brazil, Canada, China, Columbia, Czech Republic, Germany, Denmark, Ecuador, Spain, France, Great Britain, Greece, Hungary, India, Ireland, Italy, South Korea, Luxemburg, Mexico, the Netherlands, New Zealand, Peru, Russian Federation, Sweden, Slovenia, Taiwan and Ukraine.
- Recall number
- Z-0482-2011
- Recall status
- Terminated
- CDRH record ID
- 91735
- FDA classified
- 2010-11-26
- Recall event number
- 55760