Z-0482-2021: ViewRay MRIdian Linac System: Model No. 20000, CE, Rated Supply Voltage - 380/480 VAC, 3 - Frequency Ran

Recall record Z-0482-2021 for Viewray, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0482-2021
Company in FDA record
Viewray, Inc.
FDA record creation date
Nov 20, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential that components of the receive coil can reach elevated temperatures which has the potential to cause a burn injury to the patient or user.

Source facts

Action
On 10/20/2020, the firm, ViewRay, Inc, emailed an "Urgent Device Correction" notification to all affected consignees. In addition to informing consignees about the correction, the firm asked consignees to take the following actions.: 1. For Customer using TPDS software ver. 5.2.5 or earlier with optional 3 Plane imaging: - Do: Maintain a distance of at least 5 cm (~2) between the receive coil and the inner surface of the bore wall when using 3 Plane cin treatment imaging (Figure 3 & 4). - Do Not: Select 3 Plane cin treatment imaging if a distance of at least 5 cm (~2) cannot be maintained between the receive coil and the inner surface of the bore wall during treatment. 2. For Customer Using TPDS software ver. 5.3 or later with optional 8 fps enabled: - Do Not: Use 8 fps cin treatment imaging until further guidance is provided by ViewRay. - Do: Use 4 fps cin treatment imaging when treating a patient. 3. If you have any questions related to this letter, please contact ViewRay Customer Support at support@viewray.com or call +1 855-286-8875.
Initiated
2020-10-20
Root cause
Device Design
Terminated
2023-10-13
Distribution
Worldwide Distribution: US (nationwide) to states of: CA, NY, FL, IL, MI, MO, OR, WI and OUS countries of: France, Germany, Italy, United Kingdom, Taiwan, Netherlands, Israel, and Denmark.
Recall number
Z-0482-2021
Recall status
Terminated
CDRH record ID
184271
FDA record created
2020-11-20
Recall event number
86661

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗