An event has been reported to Varian which entails excessive connector resistance, which caused the actual jaw positions to differ from the intended jaw positions without warning the operator.
Source facts
Action
Varian Medical Systems sent an Urgent Medical Device Correction - Urgent Field Safety Notice dated November 28, 2011 ***Expanded to include product #2 on December 16, 2011***, to all affected customers.
***12-12-12, AMENDED INFO ***: An additional letter was sent to consignees on 11/5/12 that Varian identified as also being affected by the recall.***
The letter identified the product, the problem, and the action to be taken by the customer.
Customers were instructed to establish a comprehensive quality assurance program before beginning treatments for the day, quality assurance tests must be performed.
Additionally, any discrepancies in imaging field sizes vs planned field size should be brought to the attention of the medical physicist for investigation before treatment continues.
Customers were asked to complete the attached Proof of Notification or Receipt Verification Card once they have read the document and return it to Varian Medical Systems.
For question regarding this recall:
USA and Canada 1-888-827-4265
Europe +41 41 749 8844
Initiated
2011-11-28
Root cause
Device Design
Terminated
2015-06-02
Distribution
Worldwide Distribution
Recall number
Z-0483-2012
Recall status
Terminated
CDRH record ID
106143
FDA classified
2012-01-11
Recall event number
60659
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.