Open original FDA record ↗Recall record
- FDA identifier
- Z-0483-2021
- Company in FDA record
- Bayer Medical Care, Inc.
- FDA record creation date
- Nov 23, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Specific lots of MEDRAD Stellant FLEX 150 mL syringe kits may have holes and/or cracks that have the potential to allow fluid to leak out of the syringe.
Source facts
- Action
- On 10/08/020, Bayer sent all of their affected customers an Urgent Medical Device Field Corrective Action (recall) letter via e-mail. The recall letter requested that all of the affected customers immediately take the following steps: 1. Review your current inventory. Quarantine any affected batch/lot numbers referenced in the below chart. Instructions are included in this letter on how to identify batch/lot. 2. Complete the included response form and submit through email to CLSCRiCustomerLogistics@bayer.com or contact the Customer Logistics team at 724-940-7890. Please complete the Customer Response Form regardless of whether you have affected product or not. This action will assist us in tracking all affected product. 3. You will receive a Return Goods Authorization (RGA) number to return any affected Syringe kits remaining in your inventory for a credit. We can send new syringe kits with a separate purchase order as necessary. If product is needed immediately to avoid a stock out, expedited shipping options will be available. For expedited shipping options, please contact the Customer Logistics team at 724-940-7890. 4. Place the affected product in a box and label the outside of the container with the RGA number in large, bold writing. Use the shipping label provided via email when you receive the RGA number. 5. Upon Bayer receipt of return product, you will receive a credit on your account. Please distribute this letter to other appropriate departments and personnel within your institution who may have inventory that may be affected by this field corrective action.
- Initiated
- 2020-10-05
- Root cause
- Process control
- Terminated
- 2021-03-19
- Distribution
- Distributed to healthcare facilities nationwide throughout the US.
- Recall number
- Z-0483-2021
- Recall status
- Terminated
- CDRH record ID
- 184202
- FDA record created
- 2020-11-23
- Recall event number
- 86619