Z-0483-2025: NovoHip Biolox Ball Head 32mmL Ref: 1345-2803 NovoHip Total Hip System - NovoHip Total Hip System is a n

Recall record Z-0483-2025 for Novosource, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0483-2025
Company in FDA record
Novosource, LLC
FDA classification date
Nov 18, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Mislabeled: Packaged in a box labeled as 32 long but containing a 28 long head by physical marking, may result in joint instability if implanted

Source facts

Action
NovoSource LLC issued Urgent Medical Device Recall letter on 9/13/24 via email. Letter states reason for recall, health risk and action to take: "Immediate Action Discontinue Use: All customers and healthcare providers should immediately discontinue the use of the affected product batches. Surgical teams should not use any Biolox Delta Ceramic Head implants from the mislabeling-affected batches in any hip replacement surgeries. "Return or Segregation of Affected Product: Affected implants should be removed from active inventory and returned to NovoSource. Providers should follow instructions for proper packaging and shipping of the returned products. Any products not yet returned should be clearly labeled as "Do Not Use" and quarantined from inventory until shipment back to NovoSource. Instructions on how to return products will be provided, including the appropriate return address and contact person for support during this process. " For Unused Products: Customers should ensure that all unused products from the affected batches are returned to NovoSource. This will prevent any risk of improper head-liner fitting due to the mislabeling. "For Implanted Products: If the mislabeled product has already been implanted, surgeons should be prepared to monitor the patient for symptoms consistent with the mismatch. A mismatch of a 32 long head implanted into a 28 liner would likely have instability issues. A 28 long head implanted into a 32 liner will likely act very similarly in stability and alignment, while wear should not be catastrophic. The surgeon should monitor the patient to determine if symptoms appear that would result in the need to revise. Actions to Be Taken Pending Corrective or Removal Action: "Stop Usage: Healthcare facilities should stop any further use of the affected product batches pending completion of the recall process. NovoSource will provide updates on the investigation and resolution. "Notification to Surgical Teams: IF IMPLANTED, it is essential
Initiated
2024-09-19
Root cause
Mixed-up of materials/components
Distribution
Florida, West Virginia
Recall number
Z-0483-2025
Recall status
Open, Classified
CDRH record ID
210738
FDA classified
2024-11-18
Recall event number
95598

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗