Reason stated in the source
This firm has received Product Complaint Reports concerning bending of the pivot plate. The elevator base of the CosyCot Infant Warmer incorporates a pivot plate which forms part of the mechanism allowing the CosyCot to raise and lower in height. A bent pivot plate generally results in the Warmer being unable to elevate properly. Other component failures may result.
Source facts
- Action
- Fisher & Paykel Healthcare commenced sending out Product Correction Notifications letter via U.S. Postal Service certified mail October 9, 2006. This Product Correction Notice explained the nature of the problem and include a reminder of the maximum recommended weight of accessories. Two FAX Response Forms were provided, Part A for initial receipt of the Product Correction Notice and Part B for confirmation of the completed upgrade.
- Initiated
- 2006-10-09
- Root cause
- Other
- Terminated
- 2012-04-26
- Distribution
- Nationwide and Canada
- Recall number
- Z-0485-2007
- Recall status
- Terminated
- CDRH record ID
- 49741
- FDA classified
- 2007-01-11
- Recall event number
- 37003