Z-0485-2021: MEVION S250i, MEVION S250; Proton Radiation Treatment System - Product Usage: intended to deliver proton

Recall record Z-0485-2021 for Mevion Medical Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0485-2021
Company in FDA record
Mevion Medical Systems, Inc.
FDA record creation date
Nov 24, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Treatment beam information disappears on Treatment Console screen while beam delivery continues potential for harm to a patient could occur

Source facts

Action
Mevion issued User Notice by mail or FedEx, with return receipt confirmation on October 22, 2020. The letter states reason for recall, health risk and action to take: The User should actively monitor the TC screen during beam delivery. In the event the TC screen beam information appears blank during Beam On, the user should: Immediately suspend treatment by selecting Beam Off on the Beam Control Box. Record the MU counts displayed on the HMUC before any other action is taken. Consult physics to make manual adjustments of the plan. Notify Field Service to retrieve machine records of the partial beam delivery. Resumption of treatment requires Setup workflow to be repeated. Corrective Action All users should be notified of this scenario and the proper response. Please contact Mevion Customer service with any questions or concerns. Please distribute this notice to all users of the Mevion Medical Systems device at your organization who may be affected by this issue User confirmation of receipt shall also be received from each affected customer site. Customer Service Engineers are also permanently in working residence at customer sites and will ensure awareness of the issue.
Initiated
2020-10-19
Root cause
Device Design
Terminated
2021-08-31
Distribution
Worldwide distribution - US Nationwide distribution including in the states of DC, FL, MO, NJ, OK and the country of Netherlands.
Recall number
Z-0485-2021
Recall status
Terminated
CDRH record ID
184325
FDA record created
2020-11-24
Recall event number
86693

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗