Z-0487-2026: Affinity NT Oxygenator, Oxygenator with Balance Biosurface and Uncoated Cardiotomy/Venous Reservoir, Prod

Recall record Z-0487-2026 for Medtronic Perfusion Systems. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0487-2026
Company in FDA record
Medtronic Perfusion Systems
FDA classification date
Nov 10, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Medtronic identified an Affinity NT Oxygenator unit with decreased CO2 gas transfer rates that did not meet performance requirements.

Source facts

Action
An Urgent: Medical Device Recall notification letter dated September 2025 was sent to customers. Patient Recommendations: Patient previously supported with the impacted device face no additional risk from the issue described in this communication and should continue to be monitored per your practice s normal follow-up procedures. Customer Actions: Medtronic requests that you take the following actions: Review your inventory for listed lot number. If your inventory contains a product with Lot Number 231621003, please refer to the NT Affinity Oxygenator to identify the Serial Number. For guidance, consult Attachment A Identifying Affected Product. If you have the unit with Serial Number 5117755016 immediately quarantine and return it to Medtronic by contacting Customer Service at 1-800-854-3570, Option 1 then Option 4, and referencing this communication to initiate return and exchange or credit of this unit. Your Medtronic sales representative can assist you in the return of unit as necessary. Complete the enclosed Customer Confirmation Form and email to RS.CFQFCA@medtronic.com. This form must be returned even if you do not have the affected product in your possession. Please maintain a copy of this communication in your records. Medtronic: Report complaints at 1-800-854-3570, option 4, or email to rs.structuralheartfieldassurance@medtronic.com. If you have any questions regarding this communication, please contact your Medtronic Field Representative.
Initiated
2025-09-12
Root cause
Under Investigation by firm
Distribution
US States: AZ, CA, TN
Recall number
Z-0487-2026
Recall status
Open, Classified
CDRH record ID
216369
FDA classified
2025-11-10
Recall event number
97702

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗