Skip to content

Public FDA record

Z-0488-05: Deltec branded Administration Sets for use with Deltec 3000 and 3100; Graseby 500, 505 and AVI Volumetric

Recall record Z-0488-05 for Smiths Medical MD, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0488-05
Company in FDA record
Smiths Medical MD, Inc.
FDA classification date
Feb 2, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Administration Sets which are used with selected infusion pump systems may have a defect within the fluid path above the upper chamber of the set cassette which does not allow the pump valve to close completely and can allow fluid to leask past and may cause an inaccurate infusion rate..

Source facts

Action
Customers were sent a product Safety & Recall notification via email or regular mail begining 12/03/04 through 12/08/04. A Product Recall form requested the customer for feedback as to Quantity Used, Quantity to be Retained and Quantity to be returned . The Customer Recall Notication states that if no Administration Sets from alternative lots are available and the use of the volumetric pump is essential, follow the instructions in the attached Customer Information Bulletin, which describes how to identify a set which does not have the defect. Customers who find product subject to the recall are to contact Smith Medical Customer Service Department to make arrangements to have the affected products returned to Smiths.
Initiated
2004-12-03
Root cause
Other
Terminated
2006-06-25
Distribution
worldwide.
Recall number
Z-0488-05
Recall status
Terminated
CDRH record ID
36439
FDA classified
2005-02-02
Recall event number
30603

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.