A safety problem with installation and maintenance could lead to the breakage of the two steel cables that support the equipment to its roof anchor. Due to this, the equipment may fall and cause harm to the patient, user, or third parties.
Source facts
Action
Initial communication from SEDECAL to all our distributors was performed on 07/18/2018 by email. The letter explained the problem and requested the consignee cease use of the unit if they observe an abnormal displacement in the vertical axis or any suspicious noise during the vertical movement.
Initiated
2018-07-18
Root cause
Device Design
Distribution
US and Canada
Recall number
Z-0488-2019
Recall status
Open, Classified
CDRH record ID
168412
FDA record created
2018-11-20
Recall event number
81377