Skip to content

Public FDA record

Z-0489-2010: LACTOSORB SYSTEM 2.0MM ADJUSTABLE SELF-DRILLING TAP, REF: 915-2075, NON-STERILE PRODUCT, BIOMET MICROFIX

Recall record Z-0489-2010 for Biomet Microfixation, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0489-2010
Company in FDA record
Biomet Microfixation, Inc.
FDA classification date
Jan 4, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Biomet Microfixation is recalling LactoSorb Self Drilling Adjustable Tap Model Numbers 915-2075 Lot # 064350 and 915-2185 Lot # 885580. The two lots identified have a new style stop and an old style tap; they are not compatible, as the depth etchings do not line up correctly. Depending on the tap (1.5mm VS 2.0mm) you may drill too shallow or too deep.

Source facts

Action
Biomet Microfixation notified consignees of the recalled product beginning September 2009 via telephone and e-mail correspondence. Users were asked to return the affected product to the firm. For further information, contact Biomet Microfixation at 1-904-741-9468.
Initiated
2009-09-15
Root cause
Component change control
Terminated
2010-01-23
Distribution
USA - FL, MO, and KS; Foreign: Brazil and China.
Recall number
Z-0489-2010
Recall status
Terminated
CDRH record ID
86663
FDA classified
2010-01-04
Recall event number
53845

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗