Open original FDA record ↗Recall record
- FDA identifier
- Z-0489-2018
- Company in FDA record
- Philips North America, LLC
- FDA classification date
- Feb 8, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
Affected units may have a defective component that is intended to protect the internal circuitry from high voltages during shock delivery. The defective component may also prevent delivery of shocks in AED mode.
Source facts
- Action
- Philips Healthcare sent an Urgent Medical Device Correction Notification/Field Safety Notice letter dated June, 2017 to their customers. The letter identified the affected product, problem, hazard involved and actions to be taken. Philips is requesting that customers report if their device fails to function as intended. Philips will contact all customers for device replacement once parts are available. If you need further information or support concerning this action, please contact your local Philips representative.
- Initiated
- 2017-07-12
- Root cause
- Nonconforming Material/Component
- Terminated
- 2020-05-27
- Distribution
- No U.S. distribution, Internationally to: India, Ireland, Japan, Republic of Korea, Malaysia, Mexico, South Africa & United Kingdom
- Recall number
- Z-0489-2018
- Recall status
- Terminated
- CDRH record ID
- 159291
- FDA classified
- 2018-02-08
- Recall event number
- 78351