Z-0489-2019: Minicon Clinical Sample Concentrators for In Vitro Diagnostic Use 9031 (Model: Minicon B-15) & 9051 (Mode

Recall record Z-0489-2019 for Merck Millipore Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0489-2019
Company in FDA record
Merck Millipore Ltd.
FDA record creation date
Nov 20, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

In vitro diagnostic devices, used greater than 6 months after date of manufacture, may have increased concentration times of greater than the specification of 3 hours and up to 24 hours, which may lead to delays in carrying out the subsequent analysis tests.

Source facts

Action
On May 14, 2018, the firm sent an "Urgent Field Safety Notice" to customers via email and following up with a hard-copy letter on June 6, 2018 via United States Postal Services. The letter described the product, problem and actions to be taken. The corrective action being taken by the firm is to reduce the shelf life of the device from 12 months to 6 months. The customers are being instructed to immediately examine their inventory to determine if they have any remaining impacted product and forward a copy of this letter to the personnel in your organization that utilize this product and/or need to be made aware of this issue. In addition, if the customer has further distributed this product, please identify its customers and notify them at once of this issue. The notification to their customers should include a copy of this notification and attachment 2 Customers are instructed to please stop use of these lots if older than 6 months and destroy any remaining inventory older than 6 months, and to complete and return the receipt confirmation form to QACorkGroup@merckgroup.com. If you would like more information regarding this notification, please contact us at QACorkGroup@merckgroup.com or go to our website to find the phone number of your local office at www.emdmillipore.com/offices.
Initiated
2018-05-14
Root cause
Process control
Terminated
2021-08-05
Distribution
Worldwide Distribution: US (nationwide) to states of: AZ, CA, CT, DC, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NJ, NY, OH, OK,OR, PA, RI, TN, TX, VA, VT, WA and WI; and countries of: Canada and Australia.
Recall number
Z-0489-2019
Recall status
Terminated
CDRH record ID
167486
FDA record created
2018-11-20
Recall event number
81122

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗