Z-0489-2021: Box Label: ION Fully Articulating Catheter, REF 490105, VER 08 - Product Usage: used to navigate the Ion

Recall record Z-0489-2021 for Intuitive Surgical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0489-2021
Company in FDA record
Intuitive Surgical, Inc.
FDA record creation date
Nov 25, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The distal tip ring of the fully articulating catheter may become dislodged during the procedure.

Source facts

Action
On 10/23/2020, the firm sent an "Urgent Medical Device Correction" Notification to customers via a courier service informing them of the potential for the distal tip ring to become dislodged. All lots of the affected product (VER: 08) are potentially impacted. The Recalling Firm is instructing customers to: 1.) Prior to start of the procedure, with the Vision Probe fully inserted and connected to the back end of the Catheter, ensure that the Vision Probe does not extend beyond the Catheter. a. If the Vision Probe extends past the Catheter do not use the Catheter in a procedure. 2.) At the end of the procedure, confirm that the distal tip ring is still intact (with the Vision Probe fully inserted and connected to the backend of the Catheter, confirm that the Vision Probe does not extend beyond the Catheter. a. If the Vision Probe extends past the Catheter, perform fluoroscopy to ensure that the Catheter distal tip ring does not remain in the patient. In addition, please follow the instructions outlined in the Ion System, Instrument and Accessories User Manual (PN 553990), specifically: 1.) Inspect all instruments and accessories for potential damage prior to clinical use. 2.) Use only compatible Cook Captura biopsy forceps. Failure to use compatible forceps may result in stuck forceps or Catheter damage. a. Ensure that the Cook Captura biopsy forceps are fully closed when withdrawing into the Catheter after biopsy. b. Slowly insert and retract (do not use excessive force or rapidly withdraw) the Cook Captura biopsy forceps into the Catheter. If the Catheter distal tip ring is missing at any point during inspection, contact Customer Service immediately and return the Catheter and the distal tip ring, if available, through the standard RMA process. In addition, the Recalling Firm is asking customers to: 1. Inform necessary hospital personnel about this issue 2. Complete the attached Acknowledgement Form and return via email Recalls@intusurg.com Att
Initiated
2020-10-23
Root cause
Device Design
Terminated
2022-12-05
Distribution
US Nationwide distribution including in the states of CA, CO, DC, FL, IA, IL, MA, MD, MI, MN, NC, NJ, NY, OH, PA, TX, UT, and VA.
Recall number
Z-0489-2021
Recall status
Terminated
CDRH record ID
184275
FDA record created
2020-11-25
Recall event number
86664

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗