- Z-0489-2026
- Company in FDA record
- Abbott Laboratories
- FDA classification date
- Nov 12, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Software issue for hq analyzer results in system not visibly applying appropriate flagging to results when saturation is present which may result in incorrect results.
Source facts
- Action
- On September 18, 2025, Abbott Laboratories issued an "Urgent Medical Device Recall" Notification. Abbott ask consignees to take the following actions: 1. If IG/%IG is above your laboratory established reference range (e.g. greater than 2%), review the results per your laboratory procedures. A manual smear review may need to be performed to confirm the results. 2. Complete and return the Customer Reply Form. 3. If you have forwarded the product listed above to other laboratories, please inform them of this Product Correction and provide to them a copy of this letter. 4. Please retain this letter for your laboratory records. If the problem persists, contact your Abbott representative or Abbott Technical Support (1-800-727-7846) for help to determine whether the inability to communicate is related to this issue.
- Initiated
- 2025-09-18
- Root cause
- Software design
- Distribution
- Worldwide distribution: US (nationwide) to states of: AL, CA, CO, FL, IN, MN, NC, NY, OK and OUS (foreign) to countries of: ALBANIA, ARGENTINA, AUSTRALIA, BANGLADESH, BELGIUM, BOSNIA AND HERZEGOVINA, BRAZIL, BULGARIA, CANADA, CAYMAN ISLANDS, CHILE, COLOMBIA, CROATIA, CZECH REPUBLIC, DENMARK, EGYPT, FRANCE, GERMANY, GREECE, HONG KONG, HUNGARY, INDIA, INDONESIA, ITALY, JAMAICA, JAPAN, JORDAN, KUWAIT, LEBANON, LITHUANIA, MALAYSIA, MALDIVES, MEXICO, NETHERLANDS, NORWAY, OMAN, PAKISTAN, PANAMA, PHILIPPINES, POLAND, PORTUGAL, QATAR, ROMANIA, SAUDI ARABIA, SERBIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWITZERLAND, THAILAND, TRINIDAD & TOBAGO, TURKEY, UAE,UNITED KINGDOM, URUGUAY, and VIETNAM.
- Recall number
- Z-0489-2026
- Recall status
- Open, Classified
- CDRH record ID
- 216573
- FDA classified
- 2025-11-12
- Recall event number
- 97788