Strains of Acinetobacter radioresistens cross-react with both E. coli/P. aeruginosa PNA FISH and EK/P. aeruginosa PNA FISH to produce a false positive red signal.
Source facts
Action
AdvanDx issued a customer letter dated October 6, 2010 with a PDF file of a new package insert and an acknowledgement form to fill out and return to AdvanDx. Customers were instructed to replace the package insert in any existing inventory with the enclosed package insert.
Customers can contact AdvanDx technical support about this action at 1-866-376-0009.
Initiated
2010-10-06
Root cause
Device Design
Terminated
2011-06-06
Distribution
Worldwide Distribuition: Throughtout the US, and to the countries of Denmark, Italy, Sweden, and UK.
Recall number
Z-0490-2011
Recall status
Terminated
CDRH record ID
95159
FDA classified
2010-11-29
Recall event number
57002