Z-0490-2024: STAT-Site M BHB Test Strips, used with STAT-SITE M Beta-HB Meter as an in vitro diagnostic for blood keto

Recall record Z-0490-2024 for Stanbio Laboratory, LP. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0490-2024
Company in FDA record
Stanbio Laboratory, LP
FDA classification date
Dec 6, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Deterioration in the stability of the Test Strips results in diminished low-range detection capability and inability to meet linearity performance claims. The root cause of the stability deterioration in is identified as the change in material that caused a pH shift in the coating solution resulting in precipitation during coating runs. The device defect can cause false negative results that may lead to a delay in treatment of patients with associated conditions such as diabetic ketoacidosis.

Source facts

Action
On 10/2/23, recall notices were sent to customers who were asked to do the following: 1) Discontinue use, quarantine and dispose of any remaining inventory of the identified affected product/lot(s). 2) Ensure this information is shared with your laboratory staff and other pertinent personnel. 3) If you are a distributor, complete and return the Customer Response Form and forward the recall notification to your customers/end-users without delay for completion and return. 4) If you are the end user in receipt of the recall notification from the distributor, please complete the Customer Response Form and return it to the distributor. 5) If you are an end user in receipt of the recall notification directly from the recalling firm, complete and return the Customer Response Form. 6) If you have supplied the product to another laboratory, provide them a copy of this voluntary recall notification for completion and return notification for the same action. 7) Ensure a copy of this notification is retained as part of your laboratory Quality System records. 8) Complete and return the response form to productsafety@ekfdiagnostics.com Until the investigation is finalized, if you have further questions and/or would like to discuss possible alternative products, please contact Technical Support at; 1-800-531-5535 ext. 128. Monday through Friday, 08:00 to 17:00 (CST).
Initiated
2023-10-02
Root cause
Component change control
Distribution
Worldwide - US Nationwide distribution including in the states of AZ, RI, MO, UT, AK, LA, TX, OH, GA, MN, NY, CA, NJ, IL, WA, KS, MI, NC, PA, MA, CO, KY, NM, FL, MD, VA, IN, AL, SD, MS, WV, WI, ID, MT, WY and the countries of Canada and Singapore.
Recall number
Z-0490-2024
Recall status
Open, Classified
CDRH record ID
203865
FDA classified
2023-12-06
Recall event number
93300

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗