Skip to content

Public FDA record

Z-0490-2025: a) Alcon Custom-Pak CONSTELLATION 25G PLUS CHRISTIANA CARE HLTH SERV, # 15340-11, containing TUBING, SALE

Recall record Z-0490-2025 for Alcon Research LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0490-2025
Company in FDA record
Alcon Research LLC
FDA classification date
Nov 19, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

Reported breakage in the tubing Anti-Reflux Valve (ARV) due to improper use. As a result, changes to labeling were made.

Source facts

Action
Alcon issued a letter dated October 16, 2024 "COMMUNICATION OF MEDICAL DEVICE CORRECTION FROM CARDINAL HEALTH" (ALCON CUSTOM PAK COMPONENT SUPPLIER) regarding Alcon Custom Pak surgical kits which contain Cardinal Health Salem Sump PVC Tubes - Double Lumen Nasogastric (NG) Tube. Alcon is instructing its customers to: (1) Read the attached Cardinal Health Medical Device Correction notice. Cardinal Health recommends customers review the eIFU and updated product labels. (2) Complete the attached Alcon Customer Response Form. (3) Return the Alcon Customer Response Form by Fax: 817-302-4337 Email: Market.Actions@Alcon.com at your earliest convenience. Alcon will reconcile the forms and communicate responses to Cardinal Health. No product return is requested. Please forward the attached Medical Device Correction notice to professionals within your organization who may be using Alcon Custom Pak surgical kits containing the affected Cardinal Health Salem Sump Tubing. Contact Alcon Customer Service at 1-800-862-5266 or contact your Alcon Sales Representative with questions or concerns.
Initiated
2024-10-15
Root cause
Use error
Distribution
US distribution to states of: Delaware, Illinois, and Oklahoma
Recall number
Z-0490-2025
Recall status
Open, Classified
CDRH record ID
210740
FDA classified
2024-11-19
Recall event number
95599

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗