Skip to content

Public FDA record

Z-0490-2026: HARMONIC FOCUS Shears + Adaptive Tissue Technology, 9 cm Length REF: HAR9F, Reprocessed Device for Sing

Recall record Z-0490-2026 for Stryker Sustainability Solutions. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0490-2026
Company in FDA record
Stryker Sustainability Solutions
FDA classification date
Nov 12, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to manufacturing issue, reprocessed shears, may become stuck in the "on position" resulting in continuous activation.

Source facts

Action
On September 23, 2025, Stryker issued an "Urgent: Medical Device Recall" notification dated September 22, 2025, to affected consignees. Styker asked consignees to take the following actions: 1. Inform individuals within your organization who need to be aware of this action. 2. Check your internal inventory to locate the affected product (Attachment B). Segregate the affected products, quarantine and discontinue use. 3. Please complete and return the Business Reply Form to SSSPFA@Stryker.com or use the QR Code on Attachment A. Response is required. a. If you have indicated that product remains in your inventory, a prepaid shipping label will be issued for the return of the product. Credit will be issued to your facility for products returned upon receipt. 4. Maintain awareness of this communication internally until all required actions have been completed within your facility. For any additional information or questions please reach out to SSSPFA@stryker.com or Stryker Customer Service at 800-624-4422.
Initiated
2025-09-24
Root cause
Process control
Distribution
US Nationwide distribution and OUS (foreign) to country of: Canada.
Recall number
Z-0490-2026
Recall status
Open, Classified
CDRH record ID
216355
FDA classified
2025-11-12
Recall event number
97694

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.