Open original FDA record ↗Recall record
- FDA identifier
- Z-0491-2011
- Company in FDA record
- AdvanDx, Inc.
- FDA classification date
- Dec 1, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Strains of Acinetobacter radioresistens cross-react with both E. coli/P. aeruginosa PNA FISH and EK /P. aeruginosa PNA FISH to produce a false positive red signal
Source facts
- Action
- AdvanDX sent an e-mail notification along with a PDF file to all customers on October 8, 2010, followed by written notification dated October 6, 2010. The letter identified the product, the problem, and the action the customers should take. Customers were instructed to replace the package insert(s) included with any kits in their inventory with the enclosed package insert(s). Customers were also instructed to fill out the enclosed Acknowledgement of Receipt Form and return it to AdvanDx to confirm that they received the new package insert. For questions regarding this recall contact AdvanDx Technical Support at 1-888-376-0009.
- Initiated
- 2010-10-08
- Root cause
- Device Design
- Terminated
- 2011-06-06
- Distribution
- Worldwide Distribution - USA including IL, MI, NY, and PA and the country of Denmark
- Recall number
- Z-0491-2011
- Recall status
- Terminated
- CDRH record ID
- 95145
- FDA classified
- 2010-12-01
- Recall event number
- 56997