Open original FDA record ↗Recall record
- FDA identifier
- Z-0491-2014
- Company in FDA record
- Carestream Health Inc.
- FDA classification date
- Jan 3, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
It was discovered that Carestream received a report from a customer that 2 different patients were re-exposed due to the fact that the processed images were not able to be found in the system.
Source facts
- Action
- Carestream sent a letter to all affected customers informing them of the issue, how to detect the issue and actions to be taken by Carestream to correct the issue. The actions to be taken by Carestreamn to correct the issue will include the installation of a software patch to mitigate the effects of the observed failure and remove the possibility of a re-exposure in this failure mode. The correction will be implemented within the US to all affected devices at identified customer sites. For further questions call Carestream Customer Care at 1-800-328-2910, 8 am through 8 pm Eastern Standard Time.
- Initiated
- 2013-08-29
- Root cause
- Radiation Control for Health and Safety Act
- Terminated
- 2015-11-18
- Distribution
- Nationwide
- Recall number
- Z-0491-2014
- Recall status
- Terminated
- CDRH record ID
- 123275
- FDA classified
- 2014-01-03
- Recall event number
- 66744