This recall has been initiated due to a potential malfunction of the photo interrupter (Pl). The potential malfunction has been attributed to aging and premature deterioration of the Pl collector current.
Source facts
Action
Customers were sent a recall notification by email on July 11, 2017 informing them that their local Field Service Engineer would contact them to make arrangements to proactively replace the affected parts. Follow up notifications will be by Federal Express with verification of delivery.
Initiated
2017-06-11
Root cause
Nonconforming Material/Component
Terminated
2019-12-31
Distribution
Worldwide Distribution - U.S. Nationwide - Latin America
Recall number
Z-0491-2018
Recall status
Terminated
CDRH record ID
160396
FDA record created
2018-01-31
Recall event number
78776