Open original FDA record ↗Recall record
- FDA identifier
- Z-0491-2022
- Company in FDA record
- Medtronic Perfusion Systems
- FDA record creation date
- Jan 12, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Product is labeled as non-pyrogenic but endotoxin testing was not performed.
Source facts
- Action
- Medtronic began notifying customers on 10/18/2021, via 2-day UPS delivery, with letter titled "URGENT MEDICAL DEVICE RECALL." Customers were instructed to review inventory for affected Medtronic Perfusion Tubing Packs, identify and quarantine any affected product found in inventory, return any affected product on hand by contacting Customer Service at 1-800-854-3570, Option1 then Option 4, and referencing this communication to initiate a return and credit of unused product. Additionally, customers were instructed to complete the enclosed Customer Confirmation Certificate and scan and email to RS.CFQFCA@medtronic.com. This form must be returned even if there is no affected Perfusion Tubing Packs in inventory. Customers were also requested to share the notification with others in your organization as appropriate. If product within scope of this Urgent Medical Device Recall has been forwarded to another facility, please notify the facility of the Medtronic Urgent Medical Device Recall.
- Initiated
- 2021-10-12
- Root cause
- Under Investigation by firm
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of AL, AR, CA, CT, IL, IN, KY, MA, ME, MS, NY, OH, RI, TN, TX, and VA. The countries of Canada, Columbia, France, Germany, Italy, Netherlands, Saudi Arabia, Spain.
- Recall number
- Z-0491-2022
- Recall status
- Open, Classified
- CDRH record ID
- 190057
- FDA record created
- 2022-01-12
- Recall event number
- 88932