Skip to content

Public FDA record

Z-0492-2018: TosohSorter 9tray, Product Code 022931and TosohSorter 19tray, Product Code 022932. Product Usage: The AI

Recall record Z-0492-2018 for Tosoh Bioscience Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0492-2018
Company in FDA record
Tosoh Bioscience Inc
FDA record creation date
Jan 31, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

This recall has been initiated due to a potential malfunction of the photo interrupter (Pl). The potential malfunction has been attributed to aging and premature deterioration of the Pl collector current.

Source facts

Action
Customers were sent a recall notification by email on July 11, 2017 informing them that their local Field Service Engineer would contact them to make arrangements to proactively replace the affected parts. Follow up notifications will be by Federal Express with verification of delivery.
Initiated
2017-06-11
Root cause
Nonconforming Material/Component
Terminated
2019-12-31
Distribution
Worldwide Distribution - U.S. Nationwide - Latin America
Recall number
Z-0492-2018
Recall status
Terminated
CDRH record ID
160399
FDA record created
2018-01-31
Recall event number
78776

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗