Z-0492-2022: Flat Panel Detector, Model Number: TFP-800A, used with the Alphenix Systems, Item Numbers: INFX-8000V, IN

Recall record Z-0492-2022 for Canon Medical System, USA, INC.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0492-2022
Company in FDA record
Canon Medical System, USA, INC.
FDA record creation date
Jan 12, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Abnormal images such as horizontal stripes may be displayed and recovery may not not be possible by restarting the system. The worst case scenario is an operator who is working without a workable image may remove a catheter from the patient's blood vessel and damage the blood vessel.

Source facts

Action
On 12/21/2021, the firm notified affected customers via a letter titled, "URGENT: MEDICAL DEVICE CORRECTION." The letter indicated, "Corrective Action: When the new parts become available, your Canon service representative will contact you for an appointment to schedule the installation of the modified x-ray flat panel detector and will install on your system. Requests to Customers: Should this problem occur before the modified x-ray flat panel detector is installed on your system, please attempt computer reboot, and contact your Canon service representative. If you have any questions regarding this matter, please contact your Canon service representative. Please share the contents of this letter with all users and reviewing radiologist as well as clinical engineering or biomedical group at your facility. In order to comply with the FDA record keeping requirement, Canon respectfully requests that the attached form be completed and returned. Please fax it to the toll free number at the top of the form, or return via email to RAffairs@us.medical.canon. Thank you for your urgent attention to this matter." If you have any questions regarding this letter please contact the Regulatory Affairs Manager at 800.421.1968 or your Canon service representative at 800.521.1968.
Initiated
2021-12-20
Root cause
Vendor change control
Distribution
US Nationwide distribution in the states of DE, FL, LA, MT, NJ, OH, and WV.
Recall number
Z-0492-2022
Recall status
Open, Classified
CDRH record ID
190814
FDA record created
2022-01-12
Recall event number
89218

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗