Skip to content

Public FDA record

Z-0493-05: VERSAPOINT, Hysteroscopic system Connector Cable, Non-Sterile, Reusable (20 uses only). US Distribution/

Recall record Z-0493-05 for Ethicon, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0493-05
Company in FDA record
Ethicon, Inc
FDA classification date
Feb 2, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Deformation of the Verspoint Hysteroscopic System Connector Cable hand grips when autoclaved causing possible non-sterile electrodes.

Source facts

Action
The recall letter was mailed to the Director of Material Management and the Director of Operating Room on December 14, 2004.
Initiated
2004-12-16
Root cause
Nonconforming Material/Component
Terminated
2005-10-31
Distribution
Ethicon distributed the devices to 166 hospitals in the US. There are no Government sales. Product sold to other countries was distributed directly from the manufacturerin England to Ethicon distributors outside the US. The Foreign countries with accounts are: Austria, Czech Republic, Greece, Italy, Benelux, Poland, Portugal, Slovenia, Spain, Sweden, France, Germany, Ireland, U.K. Canada, China, Mexico, Taiwan, Venezuela, Korea, Australia, Brazil, and India.
Recall number
Z-0493-05
Recall status
Terminated
CDRH record ID
36436
FDA classified
2005-02-02
Recall event number
30711

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗