Open original FDA record ↗Recall record
- FDA identifier
- Z-0493-2010
- Company in FDA record
- Wright Medical Technology Inc
- FDA classification date
- Dec 8, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Titanium used in the recalled devices was discovered to have internal microstructure conditions of alloy segregation, primary alpha and/or a transformed beta matrix in some of the heat lots. The absence of alloying elements results in an area with less strength. For those instances where the product has been implanted, there is the potential increased risk of failure for these patients.
Source facts
- Action
- Wright Medical initiated the recall of the product on September 22, 2009 by contacting the distributors by telephone followed with letters addressed to the hospitals and surgeons dated October 9, 2009. At the firm, Customer Service is the contact for return instructions and replacement inventory. This phone number is 800-238-7117. Other questions regarding this matter are directed to 800-874-5630.
- Initiated
- 2009-09-22
- Root cause
- Process control
- Terminated
- 2010-12-16
- Distribution
- Nationwide, Columbia, Brazil, Portugal, UK, Belgium, Austria, and Greece
- Recall number
- Z-0493-2010
- Recall status
- Terminated
- CDRH record ID
- 85740
- FDA classified
- 2009-12-08
- Recall event number
- 53538