Open original FDA record ↗Recall record
- FDA identifier
- Z-0493-2014
- Company in FDA record
- Biomet, Inc.
- FDA classification date
- Dec 13, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Investigation identified that the outer label, inner label, and patient label all have the incorrect diameter size of 10.5mm. The screw is actually a 6.5mm diameter and correctly corresponds to the items number on the label.
Source facts
- Action
- Beginning November 25 Urgent recall electronic notices were sent by Biomet UK to all distributors with hospitals outside the USA notified by ground mail November 27,2013. The notification identified the product and reason for the recall, asking for immediate location and discontinued use of the product and return to Biomet UK, Ltd. Questions can be addressed to Biomet USA at 01 574-372-1570 M-F 8 am to 5 pm.
- Initiated
- 0013-11-26
- Root cause
- Labeling Change Control
- Terminated
- 2014-09-29
- Distribution
- Worldwide Distribution - USA Nationwide and in the countries of Europe, Italy, UK, Australia, Japan. all product in USA is under Biomet control.
- Recall number
- Z-0493-2014
- Recall status
- Terminated
- CDRH record ID
- 124345
- FDA classified
- 2013-12-13
- Recall event number
- 66967