The air bubble detection system may falsely continue to sound the air alarm after the user has disabled it.
Source facts
Action
Consignees were notified of the problem via letter dated 11/10/06 and informed that they would be contacted to schedule a service call to install software upgrades. Installation of software upgrades began at customer locations on 10/16/06.
Initiated
2006-10-16
Root cause
Other
Terminated
2009-07-07
Distribution
Worldwide distribution --- including USA and countries of Australia, Belgium, Canada, China, Dominican Republic, Egypt, Germany, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Republic of China, Singapore, South Africa, South Korea, Thailand and United Arab Emirates.
Recall number
Z-0494-2007
Recall status
Terminated
CDRH record ID
48749
FDA classified
2007-02-14
Recall event number
36498
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.