The ABX PENTRA Reagent Containers do not always sit properly into the reagent rack. In low reagent level scenarios, this could lead to insufficient reagent volume being pipetted and could lead to incorrect results being reported without an alarm.
Source facts
Action
The firm initiated their recall by letter on 11/21/2017. The letter stated the following:
"A new production of reagent container compliant for each reference is under process and will be made available very soon. In the meantime and to allow a proper operating of the analyzer ABX PENTRA 400 / PENTRA C400, it Is necessary to push the reagent containers in the reagent racks until they touch the bottom of the rack.
Please share this information with your laboratory staff, and retain this notification as part of your Quality System documentation. It is mandatory for you to complete and return the enclosed response form within 10 days so we
may maintain our records."
Initiated
2017-11-21
Root cause
Process control
Terminated
2019-11-04
Distribution
Nationwide
Recall number
Z-0494-2018
Recall status
Terminated
CDRH record ID
160140
FDA record created
2018-01-31
Recall event number
78692
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.