The LIFEPAK 12 biphasic defibrillator/ monitors may deliver inappropriate energy levels and fail to escalate energy levels when configured to do so.
Source facts
Action
The firm sent 4000 User Notification letters via FED-EX on 1/15/2005 and approximately 3000 on 1/17/2005. The User Notification letters informed consignees about the two complaints and described the firm''s findings from their investigation into the problem. The letters also requested the users check the default energy settings; report back to the firm any discrepancies found; and reset the discrepant settings which differed from the user programmed settings.
Initiated
2005-01-15
Root cause
Software design
Terminated
2006-05-06
Distribution
28,835 devices were shipped to 6874 domestic consignees.
Recall number
Z-0495-05
Recall status
Terminated
CDRH record ID
36800
FDA classified
2005-02-03
Recall event number
30915