- Recall classification
- Class II
Reason stated in the source
Internal audio cable may have diminished retention force due to damaged contacts, which could lead to intermittent or loss of audible alarms.
Source facts
- Action
- Philips sent an URGENT - Medical Device Correction letter dated November 9, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Philips is conducting a voluntary correction to replace the audio cable on affected devices. Customer were instructed to follow the "Action to be taken by CustomerUser" section of the instructions. Customers with wquestions were instructed to contact their Phililps representataive or contact the Customer Care Solutions Center at (800) 722-9377. For questions regarding this recall call 978-687-1501.
- Initiated
- 2012-11-09
- Root cause
- Component design/selection
- Distribution
- Worldwide Distribution - USA (nationwide) and Canada.
- Recall number
- Z-0495-2013
- Recall status
- Terminated
- CDRH record ID
- 114593
- FDA classified
- 2012-12-10
- Recall event number
- 63709