Open original FDA record ↗Recall record
- FDA identifier
- Z-0495-2015
- Company in FDA record
- Zimmer CAS
- FDA classification date
- Dec 16, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Internal discovery found that the MRI scans in some PSI knee guides involved an additional rotation adjustments that could potentially lead to misalignment due to a design flaw.
Source facts
- Action
- The firm, Zimmer CAS, sent an "URGENT: MEDICAL DEVICE NOTIFICATION" dated November 3, 2014 to its customers. The notice described the product, problem and actions to be taken. The customers were instructed to read this notice and be familiar with its contents; complete and return the attachment 1 CERTIFICATE OF ACKNOWLEDGEMENT form via fax to Zimmer Inc. at (574) 372-4265 or scan and email to corporatequality.postmarket@zimmer.com; and maintain a copy of the completed attachment 1 for your files. If you have questions concerning this notice, please call 514-861-4074 ext 292 or email: jean-sebastien.merette@zimmercas.com
- Initiated
- 2014-11-03
- Root cause
- Software design
- Terminated
- 2015-02-13
- Distribution
- US Distribution in states of: AZ, CO, IL, OR, and WA.
- Recall number
- Z-0495-2015
- Recall status
- Terminated
- CDRH record ID
- 131197
- FDA classified
- 2014-12-16
- Recall event number
- 69630