Software update to correct several issues that include (1)Potential data loss, (2) Study mix-up, (3) Incorrect measurements on multi-frame images, (4) Dearchiving issue, and (5) Unauthorized access of data due to inadequate permissions for shared folders.
Source facts
Action
Siemens released a software update VB20A_HF04 and a Customer Information Letter was sent with Update Instruction SY010/17/P. Service reports for the affected sites will be monitored to ensure the recall was effective.
For further questions, please call (610) 219-6461.
Initiated
2017-08-18
Root cause
Software design
Terminated
2018-08-30
Distribution
USA (nationwide) Distribution to the states of : AL, CT, FL, KS, LA, MA, MI, MN, MO, MS, NE, NY, OH, OR, PA, TX and WI.
Foreign consignees were not reported as device is manufactured outside the U.S.
Recall number
Z-0495-2018
Recall status
Terminated
CDRH record ID
161248
FDA record created
2018-01-31
Recall event number
79054