Z-0495-2021: Cardioblate CryoFlex, Surgical Ablation Probe, Model Number 60SF7, PROBE 60SF7 CRYOFLEX 7CM 26L - is a si

Recall record Z-0495-2021 for MEDTRONIC ATS MEDICAL, INC.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0495-2021
Company in FDA record
MEDTRONIC ATS MEDICAL, INC.
FDA record creation date
Nov 25, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential defect in the seals of the sterile barrier pouch used to package the CryoFlex disposables devices

Source facts

Action
In the United States (US), beginning 29-Oct-2020, a written Medical Device Correction letter will be mailed via UPS 2-day delivery to consignees who have received affected product according to Medtronic records. Consignees will be asked to complete and return a confirmation certificate, confirming they have received the notification. The letter requests the following actions: - Identify and quarantine all unused CryoFlex Surgical Ablation devices as listed in the table below. - Return all unused affected product in your inventory to Medtronic. Contact Medtronic Customer Service at 1-800-854-3570 to initiate a product return. Your local Medtronic Representative can assist you in the return of this product. - < Please complete the enclosed Customer Confirmation Form and email to RS.CFQFCA@medtronic.com. > - Please forward this notice to all those who need to be aware within your organization.
Initiated
2020-10-12
Root cause
Under Investigation by firm
Terminated
2024-08-16
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Belgium, Czech Republic, Germany, Korea.
Recall number
Z-0495-2021
Recall status
Terminated
CDRH record ID
184279
FDA record created
2020-11-25
Recall event number
86666

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗